DYNEO DD-1200FW - JULABO USA
We heat things up and cool them down.®

DYNEO DD-1200FW

Refrigerated Circulators

Part Number: 9021728

DYNEO DD-1200FW Water-cooled Benchtop Refrigerated / heating circulator. Range -50…+200 °C with 1 or 2 kW heating power and 1.24 kW cooling power at 20 °C. Adjustable centrifugal pump with 8…27 L/min flow and 1.5…10.2 PSI with diverter valve to adjust internal / external flow ratio. USB Communication port with options to add Analog or RS232.
WIth R1270 Natural Refrigerant.

Included with each unit: 2 each barbed fitting for tubing 8 and 12 mm inner dia. (pump connections M16x1 male)

If NRTL/UL CSA is required, contact sales@julabo.us

Quantity:
Part Number
9021728.N1.22
9021728.N1.22.A
9021728.N1.22.D
9021728.N1.33
9021728.N1.33.A
9021728.N1.33.D
Voltage Version
115V/60Hz
115V/60Hz
115V/60Hz
208-230V/50-60Hz
208-230V/50-60Hz
208-230V/50-60Hz
Interface
Base
Analog
Digital
Base
Analog
Digital
Heating & Cooling
Both
Both
Both
Both
Both
Both
Working Temp Range (°C)
-50...200
-50...200
-50...200
-50...200
-50...200
-50...200
PCF Pressure (psi)
1.5...10.2
1.5...10.2
1.5...10.2
1.5...10.2
1.5...10.2
1.5...10.2
PCF Rate (l/min)
8...27
8...27
8...27
8...27
8...27
8...27
Part Number
9021728.N1.22
9021728.N1.22.A
9021728.N1.22.D
9021728.N1.33
9021728.N1.33.A
9021728.N1.33.D
Working Temp Range (°C)
-50...200
-50...200
-50...200
-50...200
-50...200
-50...200
Heating Capacity (kW)
1
1
1
2
2
2
Cooling Capacity
°C
-40
-30
-20
0
20
kW
0.2
0.37
0.62
1.09
1.24
°C
-40
-30
-20
0
20
kW
0.2
0.37
0.62
1.09
1.24
°C
-40
-30
-20
0
20
kW
0.2
0.37
0.62
1.09
1.24
°C
-40
-30
-20
0
20
kW
0.2
0.37
0.62
1.09
1.24
°C
-40
-30
-20
0
20
kW
0.2
0.37
0.62
1.09
1.24
°C
-40
-30
-20
0
20
kW
0.2
0.37
0.62
1.09
1.24
Voltage Version
115V/60Hz
115V/60Hz
115V/60Hz
208-230V/50-60Hz
208-230V/50-60Hz
208-230V/50-60Hz
Power A
12A
12A
12A
11A
11A
11A
Plug Type
NEMA 5-20P
NEMA 5-20P
NEMA 5-20P
NEMA 6-20P
NEMA 6-20P
NEMA 6-20P
Pump Connection
M16x1 male
M16x1 male
M16x1 male
M16x1 male
M16x1 male
M16x1 male
Interface
Base
Analog
Digital
Base
Analog
Digital
PCF Pressure (psi)
1.5...10.2
1.5...10.2
1.5...10.2
1.5...10.2
1.5...10.2
1.5...10.2
PCF Rate (l/min)
8...27
8...27
8...27
8...27
8...27
8...27
Bath Fill Volume (L)
5...7.5
5...7.5
5...7.5
5...7.5
5...7.5
5...7.5
External Devices
Fluid Type
Water, Water/Glycol
Water, Water/Glycol
Water, Water/Glycol
water, water/glycol
water, water/glycol
water, water/glycol
Heating & Cooling
Both
Both
Both
Both
Both
Both
Dimensions W x L x H (in)
13x18.5x27.6
13x18.5x27.6
13x18.5x27.6
13x18.5x27.6
13x18.5x27.6
13x18.5x27.6
Weight (lbs)
97
97
97
97
97
97

Out of stock

Out of stock

Out of stock

Out of stock

Out of stock

Out of stock

Out of stock

Out of stock

Out of stock

Out of stock

Out of stock

Out of stock

  • DYNEO™ refrigerated/heating circulators are suitable for temperature applications with circulation to an external apparatus or immersing a device in the bath tank
  • A color TFT display with central rotary knob controller includes a multi-lingual interface, integrated programmer
  • Pt100 port for external temperature control
  • USB connection for external PC control and data-logging with digital options including RS232 port or an analog module interface for voltage or current control
  • PID temperature control provides ±0.01 °C stability
  • Infinitely variable centrifugal pump capacity of 8-27 L/min with pressure to 1.5-10.2 psi
  • Stainless steel bath tank with integrated drain port
  • Class III (FL) according to DIN 12876-1
  • R1270 natural refrigerant with 85g charge. GWP = 2, Carbon dioxide equivalent = 0.00017 t
  • Included with each unit: 2 each barbed fitting for tubing 8 and 12 mm inner dia. (pump connections M16x1 male) Water-cooled unit with G 3/4" male fittings for cooling water connections. Comes with adapters for 1/2" hose barbs.

DYNEO DD-1200FW Videos

Circulator Head Removal – CORIO & DYNEO

In this video, Electronics Service Team Leader, Nelson Hill will walk you through how to dismount a CORIO CD, CORIO CP, or DYNEO DD from a heated or refrigerated bath.